This is the multi-page printable view of this section. Click here to print.

Return to the regular view of this page.

System calibration

Automated applications of Quatm in routine system calibration and maintenance.

Overview

Besides physics experiments, Quatm also codes routine calibration tasks as standard experiment scripts, giving one-click execution and automatic recording. This covers common calibration needs such as camera dark noise, shutter response, frequency synthesiser verification and trigger-delay calibration.

Code structure

All calibration experiments follow the same structural pattern:

  • subclass the Experiment class
  • declare the device under test and the calibration parameters in build()
  • execute the calibration procedure in run() and record the results automatically

Typical calibration tasks

Camera dark-noise characterisation

Function: acquire image frames continuously under dark conditions and characterise the dark-current noise distribution.

Main steps:

  1. Switch off all light sources and set the camera exposure time
  2. Acquire N dark frames in a loop
  3. Compute the mean and the standard deviation of each pixel
  4. Save the noise map to HDF5

Devices involved: camera

Shutter response time measurement

Function: measure the delay of a mechanical shutter from the TTL trigger to fully open/closed.

Main steps:

  1. Align a photodetector with the beam behind the shutter
  2. Trigger the shutter open with a TTL signal and start high-speed acquisition synchronously
  3. Record the rise/fall edge times of the detector signal
  4. Repeat the measurement and average

Devices involved: shutter (TTL), photodetector

Frequency synthesiser verification

Function: verify the actual output frequency and power of a Valon 5009 or DDS frequency synthesiser.

Main steps:

  1. Set the frequency synthesiser to the target frequency
  2. Measure the actual output with a spectrum analyser
  3. Scan the frequency range and record the deviation
  4. Generate a calibration curve

Devices involved: Valon 5009/DDS (serial port), spectrum analyser

Camera trigger delay calibration

Function: measure the delay between the TTL trigger and the actual exposure window of the camera.

Main steps:

  1. Share the same TTL signal between the pulsed light source and the camera trigger
  2. Scan the trigger-delay parameter
  3. Detect the position at which the light pulse appears in each frame
  4. Calibrate the optimal delay value

Devices involved: camera, pulsed light source (TTL)